Portable Ventilator
MCV 200
ALLIED Healthcare Products INC
Can be operated solely on compressed gas, DC power, or AC power, offering the greatest flexibility for mass casualty responder
Simple control interface allows for just-in-time training and use by a wide range of responders
Assist-Control, external PEEP, and independent ventilation controls accommodate enhanced treatment options
Low maintenance and weatherproof design ensure low cost of ownership in stockpile or in everyday use
The MCV200 provides an additional level of preparedness when planning for an unknown
medical surge. In addition to its internal battery and compressors that provide the flexibility
of operationon AC or extended operation on DC power, the MCV200 also incorporates a completely pneumatic ventilation mode ensuring operation even when electricity is unavailable and all DC power sources are drained. The MCV200 is capable of providing ventilation using ambient air, high-pressure oxygen, or an oxygen/air mix. Like the MCV100, the MCV200
features intuitive ventilation controls and an array of adjustable alarms to enhance patient safety. The unique power source flexibility of the MCV200 combined with a rugged, weather-resistant and low-maintenance design ensures the greatest level of preparedness for mass casualty planners
Specifications - MCV200
Setting Parameters
ventilation Mode: Assist-Control
Assist-COntrol: Triggers at less than -2 cm H20
Flow Rate: 12-36 LPM
Breathing Rate: Adjustable. 0, 8-20 breaths per minute
TIdal volume: Adjustable. 200-1200 m1
Inspiratory TIme: 1 or 2 seconds
PEEP: External. 0-20 cm H20 (with PEEP adapter)
Fi02: Selectable. 21 %, 65%, 100%
Manometer Readout: Digital. 0-60 cm H20
Airway Pressure Limit: Adjustable. 5-60 cm H20
Power Characteristics
Power Source: Compressed O2, 110V-240V AC, internal battery
external12V DC battery
O2 Supply Pressure: 40-87 psi
Operating TIme on Battery
Based on average adult settings of 10 BPM, 600 m1 tidal volume
2 second inspiratory time
Air Mode (using internal compressors): Approx. 7 hours
100% O2 or 65% O2 Mode: Approx. 21 hours
Internal Battery: 12 Volt 5 Ah Rechargeable. Sealed lead acid type
Battery Recharge TIme: 5 hours (less than 24 hours when in use)
Battery Shelf Life: 3 years
Alarms
High Airway Pressure: Adjustable. 15-60 cm H20
Low Airway PressurelDisconnect: Adjustable. 0-30 cm H20
Low Source Gas: Fixed. Activates at about 40 psi (275 kPa) Temperature
Low Battery: Activates when approx. 1Y.2 hours of run time remain
Alarm Volume: Approx. 60 dB at 1 meter
Alarm Silence: Yes. 110 Seconds
Physical Characteristics
Dimensions: 3.5" x 10.3" x 14.5" (88.9 mm x 261.1 mm x 368.3 mm)
Weight: 17.2Ibs. (7.8 kg)
Temperature Ranges
Operating: -15 to 122°F (-9 to 50°C)
Storage: -40 to 140°F (-40 to 60°C)
Enclosure: ABS plastic with seals to prevent water ingress
Shock Resistance: 30" drop
Vibration Resistance: 100 G
Water Resistance: Spill and rain resistant